Probe lead length
|
1 m
|
|
Probe DC resistance
|
4.2 ±0.5 Ω
|
|
Cable characteristic impedance
|
75 Ω
|
|
Case thread
|
3/8-24 UNF-2A
|
|
Matched-system linear range
|
1.5 mm
|
|
Matched-system gap range
|
Approximately 0.25–1.75 mm
|
|
Recommended installed gap
|
1.0 mm
|
|
Matched-system incremental scale factor
|
7.87 V/mm (200 mV/mil)
|
|
Incremental scale-factor tolerance at 0–45°C
|
+12.5% / −20%, including interchangeability; measured in 0.25 mm increments
|
|
Matched-system linearity
|
<±0.06 mm
|
|
Matched-system frequency response
|
0–10 kHz, +0/−3 dB typical
|
|
Probe temperature range
|
−52 to +177°C, operating/storage
|
|
Minimum cable bend radius
|
25.4 mm
|
|
Maximum thread engagement
|
0.563 in
|
|
Recommended case torque
|
11.3 N·m
|
|
Maximum case torque
|
33.9 N·m
|
|
Maximum connector torque
|
0.56 N·m
|
|
Rated tensile strength
|
220 N at specified case/lead/connector interfaces
|
|
Tip material
|
Polyphenylene sulfide
|
|
Case material
|
AISI 304 stainless steel
|
|
Connector material
|
Gold-plated brass
|
|
Cable connector
|
Miniature coaxial ClickLoc, standard cable
|
|
Agency approval option
|
00, not required
|
|
Reference target
|
AISI 4140 steel
|
|
Flat target minimum diameter
|
8.9 mm minimum; 13 mm recommended minimum
|
|
Minimum probe-tip separation
|
25 mm for axial position; 23 mm for radial vibration
|
|
Mounting counterbore
|
9.5 mm minimum; 13 mm recommended minimum
|
|
Probe cable insulation
|
Fluoroethylene propylene (FEP)
|
|
Case torque on first three threads
|
7.5 N·m recommended; 22.6 N·m maximum
|